Rationale:
Inflammation is a universal pathological reaction and is characterized among other things by increased heat production. The question stays whether the contribution of the inflamed lung tissues to the overall exhaled breath temperature (EBT) can be reliably detected and used in everyday clinical practice.
Methods:
We have designed a simple device for assessment of EBT and explored its performance under standard indoor conditions. We made our measurements in the morning hours, documenting the ambient conditions (room temperature, humidity, atmospheric pressure), and physiological characteristics of the tested subjects (heart rate, blood pressure, otic and axillary temperature). We assessed its day-to-day reproducibility in 17 healthy volunteers and its ability to discriminate between the same subjects without respiratory disease and uncontrolled asthmatics (n=14). We also compared the EBT of the 14 asthmatics before and after anti-inflammatory treatment.
Results:
No association was found between EBT and any of the ambient conditions: room temperature, atmospheric pressure and humidity. While otic and axillary temperatures, which were measured in parallel, maintained high correlation between each other (Spearman's rho=0.71, p<0.01), EBT did not show meaningful association with any of them. The EBT (degrees C) of asthmatics (median 35.45, range 34.12-36.09) was higher than that of controls (34.84, 32.29-35.84), (p=0.009, Mann-Whitney U test). Anti-inflammatory treatment brought down the EBT of the asthmatics (34.78, 33.23-36.06), (p=0.001, Wilcoxon Signed Ranks test), while significantly improving their spirometry too.
Conclusions:
Measurements of EBT with the device we constructed are not significantly influenced by changes within the accepted range of a standard indoor environment. EBT represents a different characteristic of the human organism than otic and axillary temperatures. EBT is increased in uncontrolled asthmatics and decreases under anti-inflammatory treatment.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...