Clinical evaluation of the Welch Allyn SureBP algorithm for automated blood pressure measurement
1Plough Foundation, University of Tennessee Health Science Center, Suite 215 Washington Avenue, Memphis, Tennessee 38105, USA. bsalpert@utmem.edu
Objectives:
To clinically evaluate an inflation-based algorithm incorporated into a new automated blood pressure monitor manufactured by Welch Allyn, Inc.
Methods:
Device evaluation was performed according to the Association for the Advancement of Medical Instrumentation standard. An overabundance of patients with hypertension (32) were part of the 110 total participants. The data were also analyzed as described in the British Hypertension Society protocol.
Results:
The mean error and standard deviation for systolic blood pressure were -0.9 mmHg+/-7.2; for diastolic blood pressure -2.2 mmHg+/-6.7. These passed the Association for the Advancement of Medical Instrumentation standard requirements. By British Hypertension Society data analysis, the device achieved an AA grading. Over 90% of the cycles' blood pressure values were obtained during inflation.
Conclusions:
The SureBP inflation-based algorithm successfully passed the Association for the Advancement of Medical Instrumentation standard requirements and achieved an AA rating by British Hypertension Society data analysis. The monitor has great advantages for patient comfort and speed of readings (average 15 s per reading). As the population studied was skewed by including a much larger than needed number of patients with hypertension, clinicians can have added confidence in this new technology.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: