Workflow Based on the Combination of Isotopic Tracer Experiments to Investigate Microbial Metabolism of Multiple Nutrient Sources
Published on: January 22, 2018
Fast gas chromatography combustion isotope ratio mass spectrometry
Gavin L Sacks1, Ying Zhang, J Thomas Brenna
1Cornell University, Division of Nutritional Sciences, Savage Hall, Ithaca, New York 14853, USA.
Abstract:
We report here the first coupling of fast GC to IRMS, in a system capable of 250 ms peak widths (fwhm) at 1 mL/min flow rates, one-fifth as narrow as any previously reported GCC-IRMS system. We developed an optimized postcolumn interface that results in minimal peak broadening, using a programmable temperature vaporization injector in place of a rotary valve or backflush system to divert solvent, a narrow capillary combustion reactor followed by a cryogenic water trap with narrow-bore (<0.20 mm i.d.) transfer lines, and a narrow i.d. open split to the IRMS directly inserted into the column effluent. Quantitative combustion was demonstrated with CH4 injections. A comparison of CO2 injections with different fwhm peak widths (250, 2500, and 7500 ms) showed similar precisions, SD(delta13C)=0.2-0.3 per thousand, for injections of >600 pmol C on column; precision for the narrow peaks (250 ms) was considerably better for injections<150 pmol C on column. SD(delta13C)<1 per thousand was achievable for injections of 5-15 pmol on column for 250 ms wide peaks, 10-fold better precision than 2500 ms wide peaks, and within a factor of 3 of the counting statistics limit. For a mixture of 15 fatty acid methyl esters (FAME), 1.5 nmol C of each on column yielded typical SD(delta13Cpdb)=0.4 per thousand for fast GC and 0.5 per thousand for conventional GC. For 14 of the 15 FAME, delta13C values between the two systems were within+/-1.5 per thousand and not significantly different. Fast GCC-IRMS required one-third the run time (450 s vs 1400 s) to achieve comparable resolution. Mean peak widths for fast GCC-IRMS of the FAME were 720 ms, compared to 650 ms by fast GC with flame ionization detection. At a 15-fold dilution (100 pmol C on column for each FAME), fast GCC-IRMS achieved approximately 2-fold better precision and accuracy than similar injections on conventional GCC-IRMS. Finally, a mixture of 10 steroids (approximately 7 nmol C (100 ng) each on column) was analyzed with mean precision of SD(delta13C)=0.2 per thousand in 620 s by fast GCC-IRMS, while conventional GCC-IRMS required 1200 s and achieved poorer resolution. delta13C values for the two system were similar (Deltadelta13C
More Related Videos
08:43PTR-ToF-MS Coupled with an Automated Sampling System and Tailored Data Analysis for Food Studies: Bioprocess Monitoring, Screening and Nose-space Analysis
Published on: May 11, 2017
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview