Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner
Shinichiro Mori1, Kanae Nishizawa, Chisato Kondo
1Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Japan.
Background:
The 256-multislice CT (256MSCT) obtains volumetric data with 128-mm coverage in a single rotation. This coverage allows satisfactory visualization of the whole heart, allowing the 256MSCT to visualize the cardiac chambers and coronary arteries by cine scan without ECG gating. These characteristics provide a solution to the problems of MSCT. Although a wider beam width provides more efficient imaging over a wider coverage area, patient doses with the 256MSCT are of considerable concern.
Objective:
We assessed potential radiation exposure with the 256MSCT in a cardiac CT protocol and compared the results to those with 16- and 64MSCT (collimated 64x0.5mm using 256MSCT).
Methods:
Organ or tissue doses were measured in an anthropomorphic phantom under a coronary artery imaging protocol with the 256MSCT in cine scan mode without ECG gating, and with the 16- and 64MSCT in helical scan mode with ECG gating.
Results:
Average effective doses were 22.8mSv for the 16MSCT, 27.8mSv for the 64MSCT and 14.1mSv for the 256MSCT. The 16- and 64MSCT doses were thus approximately 1.6- and 2.0-fold higher than those of the 256MSCT.
Conclusions:
Use of the 256MSCT in cardiac volumetric cine imaging offers lower radiation exposure than 16- and 64MSCT, and suggests the potential of this equipment in single-beat cardiac imaging without ECG gating. This effective dose is acceptable for routine cardiac imaging.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...