Gunilla Björling1, Sara Axelsson, Unn-Britt Johansson
1Department of Clinical Sciences, Division of Anaesthesia and Intensive Care, Karolinska Institutet, Danderyd Hospital, National Respiratory Centre, and the Sophiahemmet University College, Stockholm, Sweden. gunilla.bjorling@sophiahemmethogskola.se
Objectives:
The objectives were to compare the duration of use of polymeric tracheostomy tubes, i.e., silicone (Si), polyvinyl chloride (PVC), and polyurethane (PU), and to determine whether surface changes in the materials could be observed after 30 days of patient use.
Methods:
Data were collected from patient and technical records for all tracheostomized patients attending the National Respiratory Center in Sweden. In the surface study, 19 patients with long-term tracheostomy were included: six with Bivona TTS Si tubes, eight with Shiley PVC tubes, and five with Trachoe Twist PU tubes. All tubes were exposed in the trachea for 30 days before being analyzed by scanning electron microscopy (SEM) and attenuated total reflectance Fourier transform infrared spectroscopy (ATR-FTIR). New tubes and tubes exposed in phosphate-buffered saline were used as reference.
Results:
Si tubes are used for longer periods of time than those made of PVC (P<.0001) and PU (P=.021). In general, all polymeric tubes were used longer than the recommended 30-day period. Eighteen of the 19 tubes exposed in patients demonstrated, in one or more areas of the tube, evident surface changes. The morphologic changes identified by SEM correlate well with the results obtained by ATR-FTIR.
Conclusions:
Si tracheostomy tubes are in general used longer than those made of PVC and PU. Most of the tubes exposed in the trachea for 30 days suffered evident surface changes, with degradation of the polymeric chains as a result.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...