Related Experiment Video
Updated: Jul 13, 2026

In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Occult intraocular foreign body: ultrasound biomicroscopy holds the key
Sushmita Kaushik1, Parul Ichhpujani, Aparna Ramasubramanian
1Advanced Eye Centre, Postgraduate Institute of Medical Education and Research, Chandigarh, 160012, India. sushmita_kaushik@yahoo.com
Purpose:
To illustrate the successful management of an occult intraocular foreign body (IOFB) lodged posterior to the iris causing persistent endophthalmitis.
Methods:
Two young male patients presented with corneal laceration, endophthalmitis and foreign body on X-ray but not localized by B-scan ultrasonography or gonioscopy.
Results:
Ultrasound Biomicroscopy (UBM) localized the foreign body lodged between the posterior iris and anterior lens surface accurately, which was successfully removed through the limbus using an external magnet.
Conclusions:
UBM is invaluable to accurately localize occult IOFBs lodged in the posterior iris or ciliary body region, which ensures good visual rehabilitation of these hitherto considered difficult cases.