Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
[A clinical study of impedance graph in verifying tracheal intubation]
Yong-xing Yao1, Zhen Jiang, Xia-hui Lu
1Department of Anesthesiology, Zhongshan Hospital, Fudan University School of Medicine, Shanghai 200032, China.
Objective:
To evaluate the clinical usefulness of impedance pneumography in determining the tube placement during endotracheal intubation.
Methods:
Thirty-six endotracheally-intubated patients for elective operations underwent general anesthesia and endotracheal intubation, and then a second identical tube was inserted into the esophagus under laryngoscopic control. The ventilation circuit was then attached either to tracheal or esophageal tube. The tube position was determined by 2 blinded examiners, one experienced and the other inexperienced, using three methods: impedance pneumography, capnography, and auscultation. The order of the tubes tested and the order of the methods used were randomized. The observation results and the time needed to determine were recorded by another assistant.
Results:
Of the 216 tests conducted, both examiners correctly diagnosed the position of the tube using impedance pneumography and capnography. In the auscultation method there were two false-negative results (with the tracheal tube identified as esophageal) and one false-positive (with the esophageal tube identified as tracheal) by the experienced examiner, while five false-negative results (with the tracheal tube identified as esophageal) and nine false-positive (with the esophageal tube identified as tracheal) by the inexperienced examiner. With the sensitivity and specificity of impedance pneumography as standards (100%), the sensitivity and specificity of the capnography were both 100% too, and the sensitivity and specificity of the auscultation method were 90% and 86% respectively, both significantly lower than those of the other 2 methods (all P<0.01). Capnography needed 3.4 s+/-1.3 s and 3.7 s+/-1.4 s to verify tracheal intubation and esophageal intubation respectively, both significantly longer than those of the auscultation methods (1.7 s+/-0.7 s and 2.5 s+/-1.7 s) and impedance pneumography (1.6+/-0.3 and 2.1+/-1.1s, all P<0.01). It took less time for impedance pneumography and auscultation to verify the tracheal intubation than to verify esophageal intubation (both P<0.01).
Conclusion:
Impedance pneumography is one of the reliable methods for diagnosing tracheal tube position.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.