Related Experiment Video
Updated: Jul 13, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Monte Carlo characterization of the M-19 high dose rate Iridium-192 brachytherapy source
1Radiation Laboratory, University of Massachusetts Lowell, 1 University Avenue, Lowell, Massachusetts 01854, USA. David_Medich@uml.edu
Abstract:
The MCNP5 Monte Carlo code was used to simulate the dosimetry of an M-19 iridium-192 high dose rate brachytherapy source in both air/vacuum and water environments with the in-air photon spectrum filtered to remove low-energy photons below delta=10 keV. Dosimetric data was organized into an away-along table and was used to derive the updated AAPM Task Group Report No. 43 (TG-43U1) parameters including S(K), D(r, theta), lamda, gL(r), F(r, theta), phi an(r), and phi an, and their respective statistical uncertainties.
Related Concept Videos
Radiation: Applications
The average...
Positron Emission Tomography