Effect of electronic compensation on plethysmographic airway resistance measurements
Simon Broughton1, Gerrard F Rafferty, Anthony D Milner
1Division of Asthma, Allergy and Lung Biology, King's College London-MRC Asthma Centre, London, United Kingdom.
Objectives:
To compare the performance of a plethysmograph which incorporated electronic compensation (Jaeger) to one which incorporated a heated humidified breathing system (Hammersmith plethysmograph).
Working Hypothesis:
The performance of a plethysmograph which incorporated electronic compensation would be impaired compared to that which incorporated a heated humidified system.
Study Design:
In vitro and in vivo comparison.
Patient Selection:
Eleven children, median postnatal age 13 (range 5-15) months.
Methods:
In vitro, the plethysmographs were assessed using known resistances (1.94, 4.85, and 6.80 kPa, equivalent to 20, 50, and 70 cm H(2)O/L/sec, respectively). In vivo, comparison was made of the results of children studied in both plethysmographs.
Results:
In vitro, the resistance results of the two plethysmographs were similar to each other and to the known resistances. In vivo, the median "effective" airways resistance result of the Jaeger (4.15 kPa/L/sec) was significantly higher than the inspiratory resistance of the Hammersmith plethysmograph (3.0 kPa/L/sec), but the median inspiratory resistances of the Jaeger were significantly lower than those of the Hammersmith plethysmograph (2.8 kPa/L/sec vs. 3.0 kPa/L/sec). The mean within patient coefficient of variability for inspiratory resistance of the Jaeger plethysmograph (16.7%) was significantly higher than that of the Hammersmith plethysmograph (11.6%) (P = 0.014).
Conclusion:
These results suggest plethysmographs which incorporate electronic compensation may be inappropriate for use in infants and very young children.
Related Concept Videos
Bode Plots Construction
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...