Evaluation of analytic and motion-resistant performance of the Mindray 9006 pulse oximeter
Liming Peng1, Cunliang Yan, Hong Lu
1Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, People's Republic of China. penglim2002@yahoo.com.cn
Background:
The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring of arterial oxygen saturation by pulse oximetry (SpO(2)) and pulse rate (PR) with the Mindray 9006 was evaluated.
Material/Methods:
Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed.
Results:
The results demonstrated that the coefficients of variation (CV) of imprecision were less than 1% for SpO(2) and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SpO(2)/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p>0.05) and between the Mindray 9006 and Masimo MS-3 (p>0.05). There was good correlation for SpO(2) between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r>0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r>0.98). The accuracy as the root-mean-square (A(rms)) was less than 3% for both SpO(2) and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO(2)/PR between the motion and rest conditions (p>0.05).
Conclusions:
The overall performance of the Mindray 9006 for SpO(2) and PR analysis is excellent and compares favorably with the Masimo MS-3.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: