Reprocessing flexible gastrointestinal endoscopes after a period of disuse: is it necessary?
A S Vergis1, D Thomson, P Pieroni
1Brandon Regional Health Centre, Brandon, Manitoba, Canada. ashleyvergis@hotmail.com
Background:
Gastrointestinal endoscopy is an integral tool in the evaluation and management of many gastrointestinal and hepatobiliary conditions. Although rare, media reports of infectious complications following gastrointestinal endoscopy persist in this new millennium. With only limited data available, society guidelines continue to suggest that endoscopes undergo a reprocessing cycle before the first patient of the day. This preliminary study aimed to assess the microbiological stability of gastrointestinal endoscopes after high-level disinfection.
Methods:
In this multiphase study, four endoscopic retrograde cholangiopancreatography (ERCP) scopes and three colonoscopes were evaluated. In phase 1, endoscopes were assayed after initial high-level disinfection and daily for a period of 2 weeks. In phase 2, this procedure was repeated to confirm phase 1 results. In phase 3, endoscopes were assayed after high-level disinfection and again following a 7-day storage period.
Results:
In phase 1, 6 of 70 (8.6 %) assays were positive. This involved 4 of 7 (57 %) endoscopes (2 colonoscopes and 2 ERCP scopes) and was limited to the first 5 days of the study. No cultures were positive in phase 2. In phase 3, one endoscope had a positive culture. Positive cultures grew only STAPHYLOCOCCUS EPIDERMIDIS, a low-virulence skin organism.
Discussion:
With proper disinfection and storage, it appears that reprocessing of gastrointestinal endoscopes is unnecessary after periods of disuse of at least 7 days and possibly up to 2 weeks. Despite recent media reports of infectious complications, society guidelines that recommend more frequent reprocessing seem to lack scientific merit and need to be revisited.
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures V: ERCP
Patient...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures II: Colonoscopy