Related Experiment Video
Updated: Jul 13, 2026

Ye's Swing Technique for Small-incision Lenticule Extraction Surgery
Published on: June 27, 2025
Comparative performance of the Zyoptix XP and Hansatome zero-compression microkeratomes
Jay S Pepose1, Susan K Feigenbaum, Mujtaba A Qazi
1Pepose Vision Institute, Missouri 63017, USA. jpepose@peposevision.com
Purpose:
To compare flap parameters produced by the Zyoptix XP and Hansatome microkeratomes (both Bausch & Lomb) and to evaluate preoperative variables contributing to flap thickness variation.
Setting:
Private practice, St. Louis, Missouri, USA.
Methods:
Flap dimensions in 75 eyes that prospectively had laser in situ keratomileusis using the Zyoptix XP microkeratome were compared with a historical control sample of 75 eyes treated with the Hansatome microkeratome. The 2 groups were matched for mean keratometry, central corneal pachymetry, spherical equivalent (SE), age, and microkeratome head size and suction ring diameter.
Results:
The mean ultrasound-measured flap thickness was 126.54 microm +/- 14.6 (SD) and 143.74 +/- 15.0 microm for the 120 microm and 140 microm Zyoptix XP heads, respectively, and 128.90 +/- 20.4 microm and 143.32 +/- 21.0 microm for the 160 microm and 180 microm Hansatome heads, respectively. The standard deviation in flap thickness was smaller and statistically significant for the Zyoptix XP (+/-14.8 microm) versus the Hansatome (+/-20.7 microm) (P = .0039, F test). Preoperative pachymetry and SE (P<.001) accounted for 20% of the variability in measured flap thickness using the Hansatome. For the Zyoptix XP, preoperative pachymetry was statistically significant in explaining the variation in flap thickness with the 120 microm head (P = .02) but not with the 140 microm head. Variation in flap thickness from either Zyoptix XP head was not statistically related to the preoperative SE.
Conclusion:
Although the 2 microkeratomes produced flaps of similar mean thickness, the Zyoptix XP showed significantly less variation in flap thickness than the Hansatome, was less affected by measurable preoperative variables such as SE, and was closer to nominal labeling.