Related Experiment Video
Updated: Jul 13, 2026

Surgical Retrieval, Isolation and In vitro Expansion of Human Anterior Cruciate Ligament-derived Cells for Tissue Engineering Applications
Published on: April 30, 2014
Broken femoral cross pin after hamstring anterior cruciate ligament reconstruction: case report
Neal C Chen1, Robert E Boykin, Peter J Millett
1Massachusetts General Hospital Hand and Upper Extremity Fellowship, Boston, Mass 02114, USA.
Abstract:
Two years after a quadrupled hamstring anterior cruciate ligament reconstruction using polylactic acid cross pin femoral fixation, a 32-year-old woman presented with symptoms of knee catching, locking, and stiffness. Diagnostic arthroscopy revealed a loose body in the anterior compartment of the knee which was determined to be part of the polylactic acid femoral fixation pin. The graft was intact and well fixed. After simple arthroscopic removal, the patient returned to full activities and resumed normal function postoperatively. We speculate that the fixation pins may have entered the notch and later degraded or fractured. Using axial magnetic resonance images, we provide preliminary data suggesting that pins angled posterior to the epicondylar axis may violate the notch. If pins are to be placed posterior to the epicondylar axis, maximum pin length can be estimated by the formula: 0.4 x the interepicondylar distance.