Computer software for the on-line measurement of the left ventricular end-systolic pressure-dimension relationship
J A Ward1, C P Kingsley, S M Silverman
1Department of Clinical Investigation, Brooke Army Medical Center, Fort Sam Houston, TX 78234-6200.
Abstract:
A microcomputer system was designed to measure the end-systolic pressure-dimension relationship (ESPDR), an index of cardiac contractility that is independent of preload, afterload, and heart rate. To test the system, pressure-dimension data were obtained from swine left ventricles and from a mathematical model of the heart. Algorithms for filtering, location of end-systole, selection of the measurement interval, and calculation of the ESPDR were evaluated on the basis of speed, precision, accuracy, and robustness. The resulting program runs on an IBM-AT and measures ESPDR on-line within 60 seconds of the start of data acquisition. By reducing the time spent in data analysis and providing rapid feedback of information, the on-line software has increased productivity and facilitated improvements in experimental technique.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...