Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
Experimental design optimization of the capillary electrophoresis separation of leucine enkephalin and its immune
Sheikh Md Enayetul Babar1, Eun Joo Song, Md Nabiul Hasan
1Bioanalysis and Biotransformation Research Center, Korea Institute of Science and Technology, Cheongryang, Seoul, Korea.
Abstract:
To optimize the capillary electrophoretic separation conditions for leucine enkephalin (LE) and the immune complex of the LE and anti-LE reaction, an analysis using a three-level, three-factorial Box-Behnken design was performed. Three separation parameters, buffer pH (X(1)), buffer concentration (X(2)), and applied voltage (X(3)), were chosen to observe the effect on separation responses. The responses were theoretical plate number, migration time of the LE peak, and resolution between the peaks. The optimum conditions and process validation were determined using statistical regression analysis and surface plot diagrams. The capillary electrophoresis optimum separation conditions were established to be 75 mM phosphate buffer at pH 7.00 with an applied separation voltage of 15 kV. By using the analysis technique, the prediction of responses was satisfactory and process verification yielded values within the +/-5% range of the predicted efficiency.
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...