Related Experiment Video
Updated: Jul 13, 2026

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Poly(methyl methacrylate) microchip device integrated with gold nanoelectrode ensemble for in-column biochemical
Kuo-Tang Liao1, Chun-Mao Chen, Hsuan-Jung Huang
1Department of Chemistry, National Sun Yat-sen University, 804 Kaohsiung, Taiwan.
Abstract:
This paper proposes a poly(methyl methacrylate) (PMMA) based microchip with an integrated gold nanoelectrode ensemble (GNEE) and a quartet-T loading channel for in-column urea/urease reactions and electrochemical detections. The on-chip GNEE electrode is fabricated using an electrodeless deposition process on a thin polycarbonate (PC) film and bonded directly onto a PMMA substrate to carry out high-performance electrochemical detections. The in-column bio-catalytic reaction of urea/urease is successfully demonstrated utilizing a novel approach based on the different electrokinetic mobilities of urea and urease in capillary electrophoresis (CE) channel. The experimental results significantly show that the GNEE electrode provides a better detection response for the reaction product of ammonia (NH(4)(+)) than a conventional planar gold electrode. The detection results demonstrate a satisfactory determination coefficient (R(2) value) and high reproducibility with a detection limit of 14.8 and 62.8 microM while detecting standard ammonia solution and the urea/urease reaction product of NH(4)(+), respectively. These results confirm the capability of the proposed device for the high-resolution CE-electrochemical detection (CE-ED) of bioanalytical reactions.