Related Experiment Video
Updated: Jul 13, 2026

Characterizing Far-infrared Laser Emissions and the Measurement of Their Frequencies
Published on: December 18, 2015
Long-term frequency stabilization of a continuous-wave tunable laser with the help of a precision wavelengthmeter
Sergey Kobtsev1, Stepan Kandrushin, Andrey Potekhin
1Laser Systems Laboratory, Novosibirsk State University, Russia. kobtsev@lab.nsu.ru
Abstract:
For the first time to our knowledge we experimentally demonstrate an efficient method for the reduction of long-term radiation line drift in single-frequency cw Ti:sapphire and dye lasers that relies on a fast and precise wavelengthmeter together with a digital-analog feedback system. Generation line drift of lasers is reduced approximately by an order of magnitude down to 40 MHz/h, which corresponds to the residual drift in readings of the wavelengthmeter itself. The implemented automatic frequency control system allows us to lock the laser generation frequency to a specified absolute value. This approach may be used in single-frequency lasers of different types (solid-state, fiber, diode, dye lasers, etc.) and allows reduction by an order of magnitude or more of the long-term generation line drift in lasers that are not equipped with other systems for long-term stabilization of output radiation frequency.