Distal embolic protection during percutaneous intervention of aorto-coronary venous bypass grafts: the FIRST Trial

M Lins1, H Heuer, M Haude

  • 1University Clinic Schleswig-Holstein, Campus Kiel, Department of Cardiology, Schittenhelmstrasse 12, 24105 Kiel, Germany.

Insights

The TriActiv device effectively protected against debris during bypass graft interventions, showing a low major adverse cardiac event rate. This supports using protection systems for coronary artery bypass graft procedures.

Area of Science:

  • Cardiovascular Interventions
  • Medical Device Technology
  • Interventional Cardiology

Background:

  • Aorto-coronary venous bypass grafts (CABG) interventions risk distal embolization.
  • Debris embolization can lead to acute procedural complications.
  • The TriActiv device is a distal endovascular protection system.

Purpose of the Study:

  • To evaluate the safety and effectiveness of the TriActiv device.
  • To assess the TriActiv device during CABG interventions.
  • To determine the incidence of complications during CABG interventions with TriActiv.

Main Methods:

  • A multicenter trial (FIRST) involving 195 patients across 17 German centers.
  • Patients had significant disease of a vein graft.
  • Inclusion/exclusion criteria mirrored the SAFER trial.

Main Results:

  • Acute procedural success was achieved in 98% of cases.
  • Aspirated debris was present in 96.5% of patients.
  • The 30-day rate of Major Adverse Cardiac Events (MACE) was 8.7%, with no deaths and a 7.2% Myocardial Infarction (MI) rate.

Conclusions:

  • The TriActiv system demonstrated safety and effectiveness in CABG interventions.
  • The device preserved normal post-procedural flow.
  • The low MACE rate (8.7%) supports the use of protection systems in PCI of CABG grafts.
Abstract