Related Experiment Video
Updated: Jul 13, 2026

Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and
S H Bartling1, O Majdani, R Gupta
1Department of Neuroradiology, OE 8210, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany. soenkebartling@gmx.de
Objectives:
To assess the feasibility of flat-panel detector based volumetric CT (fpVCT) scanning of the whole human skull base and maxillofacial region, which has thus far only been demonstrated on small, excised specimens. Flat-panel detectors offer more favourable imaging properties than image intensifiers. It is therefore likely that they will replace them in cone-beam CT scanners that are currently used to scan parts of the skull base and maxillofacial region. Furthermore, the resolution of current CT imaging limits diagnosis, surgical planning and intraoperative navigation within these regions. fpVCT might overcome these limitations because it offers higher resolution of high contrast structures than current CT.
Methods:
Three embalmed cadaver heads were scanned in two scanners: an experimental fpVCT that offers a scan field large enough for a whole human head, and in a current multislice CT (MSCT). 28 structures were compared.
Results:
Both scanners produced bone images of diagnostic quality. Small high contrast structures such as parts of the ossicular chain and thin bony laminas were better delineated in fpVCT than in MSCT. fpVCT of maxillofacial region and skull base was rated superior to MSCT (P=0.002) as found in this limited, experimental study.
Conclusions:
High spatial resolution fpVCT scanning of both regions in a whole human head is feasible and might be slightly superior to MSCT. fpVCT could improve diagnostic accuracy in selected cases, as well as surgical planning and intraoperative navigation accuracy.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography