Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Accuracy of combined GDx-VCC and matrix FDT in a glaucoma screening trial
Márta Tóth1, Péter Kóthy, Péter Vargha
1Department of Ophthalmology, Semmelweis University, Budapest, Hungary.
Purpose:
To assess the advantage in glaucoma screening of the use of scanning laser polarimetry with customized cornea compensation (GDx-VCC) combined with Matrix Frequency Doubling Technology (M-FDT) testing.
Methods:
In a nonpopulation-based prepublicized trial, self-recruited white participants were screened for glaucoma with GDx-VCC, with M-FDT, and by independent clinical examination. Cases with possible glaucoma as found with any of the screening methods underwent a detailed clinical investigation to verify or exclude glaucoma. Sensitivity, specificity, accuracy, likelihood ratios, and predictive values were calculated using different threshold criteria for GDx-VCC alone, M-FDT alone, and for various combinations.
Results:
Of the 233 attendees, 181 participants (345 eyes) successfully underwent the GDx-VCC and M-FDT measurements. Thirty-nine eyes of 24 participants had glaucoma (11.3% prevalence among eyes tested successfully). All but 2 of the glaucomatous eyes had only early damage. Evaluated separately, the criterion GDx-VCC NFI (normal threshold < or =30) performed best, with 97.0% specificity, 88.8% accuracy, and 25.6% sensitivity; but with only 8.5 positive likelihood ratio (PLR). For paired criteria, the best combination of GDx-VCC-screening test with M-FDT-screening test provided 99.6% specificity, 91.3% accuracy, and 28.6 PLR. For NFI combined with GDx-VCC nerve fiber bundle defect criterion, specificity was 99.0%, accuracy 89.6%, and PLR 18.0. However, the sensitivities in the 2 cases fell to 12.0% and 18.0%. For a triple combination of M-FDT-screening test with the latter pair of criteria, sensitivity increased to 41.7% and PLR (13.6) still remained clinically useful.
Conclusions:
In a self-recruited white population with relatively high risk for mild glaucomatous damage, a combination of GDx-VCC together with M-FDT could usefully be employed for mass glaucoma screening.