Some aspects on the use of the Biotest RCS Air Sampler in unidirectional air flow testing
1Royal Institute of Technology, Stockholm, Sweden.
Abstract:
The Biotest Air Sampler RCS is often used for monitoring microbiological quality of air. Unidirectional air flow tests show that the dispersion region of contaminants around the RCS sampler is much larger than that of undisturbed parallel air flow. This must be taken into consideration when the RCS air sampler is used within pharmaceutical production, especially during aseptic operations.