Measuring fetal and maternal temperature differentials: a probe for clinical use during labour

N J Randall1, K Bond, J Macaulay

  • 1Department of Obstetrics and Gynaecology, Charing Cross and Westminster Medical School, West London Hospital, UK.

Journal of Biomedical Engineering
|November 1, 1991
PubMed

The healthy fetus maintains a higher temperature than that of its mother during gestation and labour. This results from the thermal balance between the heat generated by the fetus and the heat loss to its maternal surroundings. The heat loss can be by heat exchange via blood flowing in the umbilical cord and placenta, and via conduction through the fetal skin and amniotic fluid to the maternal wall. The temperature difference between the fetal and maternal tissue may reflect the metabolic state of the fetus and the magnitude and changing patterns of placental blood flow during labour. Physiological changes, such as those induced by epidural analgesia, and fetal infection have been shown to exhibit an increase in the absolute temperature. An intrauterine probe, previously used for non-invasive ECG detection, has been equipped with temperature sensors that measure fetal and maternal skin temperature in utero. Laboratory tests to characterize the performance of the probe reveal that absolute and differential temperatures can be resolved to around 0.01 degree C with a thermal time constant of approximately 9 s. Ideally the probe body should have infinite thermal insulation or thermal shunting across the probe will occur reducing the measured temperature difference. In this initial probe design, a high thermal isolation between sensors has been achieved but is not perfect, resulting in around 85% of the actual temperature difference across the probe being registered. Average feto-maternal differences of 0.2 degrees C have been measured in a clinical investigation.

Related Concept Videos

Equipments Used to Measure Body Temperature01:13

Equipments Used to Measure Body Temperature

Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites01:14

Temperature Measurement Sites

A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane01:14

Assessing Body Temperature - Tympanic membrane

Assessing tympanic membrane temperature involves using a tympanic membrane thermometer (TMT). Here is a step-by-step guide:
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal01:27

Assessing Body Temperature - Rectal

Rectal temperature measurement is considered the most precise method for assessing core body temperature and typically registers higher than oral temperature. For adults, the rectal thermometer should be inserted 1 to 1.5 inches into the rectum to obtain the most accurate reading.
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla01:14

Assessing Body Temperature - Axilla

Procedural Guide for Assessing Axillary Body Temperature using a Digital Thermometer:
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery01:19

Assessing Body Temperature - Temporal Artery

Here is a stepwise guide to assessing the body temperature at the temporal artery using a temporal artery thermometer
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.  
Step 3: Assess the patient's forehead...