Related Experiment Video
Updated: Jul 13, 2026

Isolated Lung Perfusion System in the Rabbit Model
Published on: July 15, 2021
Optional active compliance chamber performance in a pulmonary artery-pulmonary artery configured paracorporeal
Scott K Alpard1, Dongfang Wang, Donald J Deyo
1Department of Surgery University of Texas Medical Branch, Galveston, TX 77555-0528, USA.
Introduction:
Our group has developed a paracorporeal artificial lung (PAL) attached in a pulmonary artery (PA) to PA in series configuration to address profound respiratory failure and serve as a bridge to transplant and/or recovery. We recently designed, developed and converted our passive pre-PAL compliance chamber to an active, synchronized, counterpulsating assist device to relieve right heart strain and offset increased work placed on the right ventricle when the PAL is attached. In this study, we evaluated the safety and performance of both a valved and non-valved optional active compliance chamber (OACC) in a PA-PA PAL for right heart assistance in normal adult sheep.
Methods:
Eleven sheep (30-50 kg) were divided into non-valved OACC (n = 6) and valved (n = 5) OACC groups. To mimic pulmonary hypertension, a C-clamp was placed distal to the OACC-PAL and occluded until a 20% decrease in cardiac output (CO) was achieved. The OACC was activated, and right ventricular pressure (RVP), pulmonary artery pressure (PAP), mean arterial pressure (MAP) and CO were recorded.
Results:
All eleven animals tolerated the implantation of the OACC PAL. Activation of the OACC resulted in a significant increase in CO. Systolic and diastolic right ventricular pressure decreased in both groups. Lastly, counterpulsation increased the mean PAP in all animals and peak PAP reached 89 mmHg. Despite providing right heart assistance, synchronizing the counterpulsation was technically difficult, and the high peak PA pressures resulted in anastomotic bleeding in all animals and anastomotic breakdown in 4/11 animals.
Conclusions:
An OACC PAL perfused by the right ventricle applied in series with the pulmonary circulation reduces ventricular load and improves cardiac efficiency. These preliminary data suggest the potential of an artificial lung in unloading the strained right ventricle and acting as a bridge to transplantation. The augmented peak PA pressures, resulting in bleeding and anastomotic breakdown, and complexity in synchronizing the cardiac cycle with the pulsations of the augmented OACC, compromise this configuration.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...