Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Consequences of the radial heterogeneity of the column temperature at high mobile phase velocity
Fabrice Gritti1, Georges Guiochon
1Department of Chemistry, University of Tennessee, Knoxville, TN 37996-1600, USA.
Abstract:
When a high velocity stream of mobile phase percolates through a chromatographic column, the bed cannot remain isothermal. Due to the mobile phase decompression, heat is generated along the column. Longitudinal and radial temperature gradients take place along and across its bed. The various consequences of this thermal heterogeneity are calculated and their effects on the column efficiency investigated for a 0.46 cm x 25 cm stainless steel column packed with 5 microm particles. The maximum pressure drop applied was varied from 0.1 to 2 kbar. The amplitude of the longitudinal temperature gradient can be estimated on the basis of the integral heat balance equation applied to the whole column and of measurements of the eluent temperature at the column exit. Assuming that the radial gradient is parabolic and the longitudinal gradient linear, the amplitude of the radial gradient can be determined on the basis of the energy balance across the column and of direct measurements of the radial gradient at high inlet pressures. A radial temperature gradient causes a radial distribution of the eluent viscosity, hence of its local velocity. The result is that bands move faster in their center than along the wall, become warped, hence a radial concentration gradient, similar in origin to the one observed in open cylindrical tubes. Diffusion relaxes this gradient. If there is only a longitudinal temperature gradient, the column efficiency would be 30% smaller for a 2 kbar pressure drop than if there is no longitudinal temperature gradient. However, when both a longitudinal and a radial temperature gradient coexist, there is a large loss of efficiency. If the influence of the diffusive relaxation of the radial concentration gradient is neglected, the peak shape would be broad and exhibit a marked shoulder.
Related Concept Videos
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Distribution of Molecular Speeds
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Steady, Laminar Flow in Circular Tubes