Quantitative Analysis of Dietary Vitamin A Metabolites in Murine Ocular and Non-Ocular Tissues Using High-Performance Liquid Chromatography
Published on: December 27, 2024
Optimization and validation of a high performance liquid chromatography method for the simultaneous determination of
Lubor Urbánek1, Dagmar Solichová, Bohuslav Melichar
1Charles University in Prague, Department of Analytical Chemistry, Faculty of Pharmacy, Heyrovského 1203, 500 05 Hradec Králové, Czech Republic.
Abstract:
In this study a novel, simple and rapid reversed-phase high performance liquid chromatography (HPLC) procedure for simultaneous determination of vitamins A and E (retinol and alpha-tocopherol) in blood serum has been developed and validated using monolithic column and diode-array detection (DAD). The monolithic column Chromolith Performance RP-18e (100 mm x 4.6 mm) was operated at ambient temperature. One hundred percent methanol at flow rate 2.5 ml min(-1) was used as a mobile phase. Detection of both compounds was performed with diode-array detector, retinol was monitored at 325 nm and alpha-tocopherol at 295 nm. The linear dependence between peak area and concentration ranged from 0.25 to 10.00 micromol l(-1) for retinol and 0.5-50.0 micromol l(-1) for alpha-tocopherol. The limit of detection (LOD) for retinol was 0.02 micromol l(-1) and limit of quantification (LOQ) was 0.07 micromol l(-1). The limit of detection (LOD) for alpha-tocopherol was 0.1 micromol l(-1) and limit of quantification (LOQ) was 0.3 micromol l(-1). Retinol was eluted in 0.8 min and alpha-tocopherol in 1.4 min. The simultaneous analysis of vitamin A and E can be achieved in less than 2 min. The implementation of monolithic column Chromolith Performance shortens the time of analysis of both vitamins four times in comparison with using traditional particulate column Pecosphere C18 (150 mm x 4.6 mm), 5 microm. This fact may play an important role for routine clinical analysis of biological samples.
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...