Related Experiment Video
Updated: Jul 12, 2026

Bridging the Bio-Electronic Interface with Biofabrication
Published on: June 6, 2012
Electrogenerated chemiluminescence biosensor based on Ru(bpy)3(2+) and dehydrogenase immobilized in
Lihua Zhang1, Zhiai Xu, Shaojun Dong
1State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Graduate School of the Chinese Academy of Sciences, Chinese Academy of Sciences, Changchun 130022, China.
Abstract:
A new electrogenerated chemiluminescence biosensor was fabricated by immobilizing ECL reagent Ru(bpy)3(2+) and alcohol dehydrogenase in sol-gel/chitosan/poly(sodium 4-styrene sulfonate) (PSS) organically modified composite material. The component PSS was used to immobilize ECL reagent Ru(bpy)3(2+) by ion-exchange, while the addition of chitosan was to prevent the cracking of conventional sol-gel-derived glasses and provide biocompatible microenvironment for alcohol dehydrogenase. Such biosensor combined enzymatic selectivity with the sensitivity of ECL detection for quantification of enzyme substrate and it was much simpler than previous double-layer design. The detection limit was 9.3 x 10(-6) M for alcohol (S/N=3) with a linear range from 2.79 x 10(-5) to 5.78 x 10(-2) M. With ECL detection, the biosensor exhibited wide linear range, high sensitivity and good stability.