Statistical design-principal component analysis optimization of a multiple response procedure using cloud point
Marcos A Bezerra1, Roy E Bruns, Sergio L C Ferreira
1Universidade Federal da Bahia, Instituto de Química, Grupo de Pesquisa em Química Analítica, Campus Universitário de Ondina, Salvador, Bahia 40170-290, Brazil.
Abstract:
A procedure has been developed for the simultaneous determination of traces amounts of Cd, Cr, Cu, Mn, Ni and Pb from saline oil-refinery effluents and digested vegetable samples using inductively coupled plasma optical emission spectrometry (ICP OES). The procedure is based on cloud point extraction (CPE) of these metals as 2-(bromo-2-pyridylazo)-5-diethyl-amino-phenol (Br-PADAP) complexes into a micellar phase of octylphenoxypolyethoxyethanol (Triton X-114). Optimization of the procedure was performed by response surface methodology (RSM) using a Doehlert design. Principal components (PC) were used to simplify the multiple response analysis. A response surface for the first PC score is useful in determining the optimum conditions for the Cd, Cr, Cu, Mn and Pb determinations whereas the second PC is highly correlated with the Ni response. Improvement factors of 22, 36, 46, 25, 65 and 39, along with limits of detection (3sigma(B)) of 0.081, 0.79, 0.38, 0.83, 0.28 and 0.69 microg L(-1), and precision expressed as relative standard deviation (%R.S.D., n=8, 20.0 microg L(-1)) of 1.5, 2.2, 3.5, 2.6, 2.5 and 2.5 were achieved for Cd, Cr, Cu, Mn, Ni and Pb, respectively. The accuracy was evaluated by spike tests in oil-refinery effluent samples and analysis of a vegetable certified reference material (NIST 1571, orchard leaves). Results found were in agreement with certified values.
Related Concept Videos
Methods of Medium Optimization
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Extraction: Advanced Methods
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.