Related Experiment Video
Updated: Jul 9, 2026

High-pressure Sapphire Cell for Phase Equilibria Measurements of CO2/Organic/Water Systems
Published on: January 24, 2014
Vapor pressure of water at its triple point: highly accurate value
Abstract:
The vapor pressure of water at its triple point was measured with greatly increased accuracy. The triple point was realized with newly designed equipment that enhanced the stability of the pressure and permitted any air released from solution to be removed by pumping. A diaphragm pressure transducer separated the water vapor from the helium used to transmit the pressure to the manometer. The pressure was measured with the National Bureau of Standards precision mercury manometer. The vapor pressure at the triple point was found to be 611.657 pascals with random uncertainties at the 99 percent confidence level of +/- 0.010 pascal. The systematic errors are estimated to be relatively insignificant.
Related Concept Videos
Vapor Pressure
Phase Diagram
Phase Diagram
Vapor Pressure of Fluid
When a liquid is placed in a closed container with a small air space, and the space is evacuated, vapor molecules will...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Static, Stagnation, Dynamic and Total Pressure