Related Concept Videos
Spectrophotometry: Introduction
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Ultraviolet–visible (UV–visible or UV–Vis) spectroscopy is an analytical technique that investigates the interaction between matter and UV–Vis light within the electromagnetic spectrum. This method is widely used for its versatility, simplicity, and relatively quick data acquisition, making it valuable for both qualitative and quantitative analysis. When UV–Vis radiation passes through a material, molecules absorb light depending on the energy required for electronic transitions. As a result...
UV–Vis Spectrometers
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
IR Spectrometers
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
Atomic Spectroscopy: Absorption, Emission, and Fluorescence
Atomic spectroscopy is a vital tool in elemental analysis, both qualitatively and quantitatively. It can be broadly divided into optical spectroscopy, mass spectroscopy, and X-ray spectroscopy methods. The optical spectroscopic methods are atomic absorption spectroscopy (AAS), atomic emission spectroscopy (AES), and atomic fluorescence spectroscopy (AFS). The first step in all three methods is atomization, where the solid, liquid, or solution-phase samples are converted into gas-phase atoms and...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Atomic absorption spectroscopy (AAS) relies on the Beer-Lambert law, which requires that the radiation source emits a narrow range of wavelengths to match the absorption characteristics of the analyte atom. The primary criteria for choosing an appropriate radiation source in AAS is to provide a precise and intense emission at specific wavelengths that will allow accurate detection of the analyte.
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...