Standardization of a multi-wavelength UV detector for liquid chromatography-based toxicological analysis
S R Binder1, A K Adams, M Regalia
1Bio-Rad Laboratories, Clinical Systems Division, Hercules, CA 94547.
Abstract:
The performance of a multi-wavelength UV detector for automated drug identification following liquid chromatographic separation was evaluated. The ability of selected wavelength ratios to distinguish two closely related drugs was considered at different concentrations. Calibration of the detector based on wavelength ratios was then utilized to standardize two different detectors and to evaluate instrument-to-instrument variation of a series of detectors. Reproducibility of the second-derivative zero intercept for these drug spectra was also evaluated. Standardization of detector performance by reference to these two parameters permitted the transfer of UV spectral libraries stored on one instrument to another without compromising the reliability of qualitative data.
Related Concept Videos
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
UV–Vis Spectrometers
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors