A Model to Simulate Clinically Relevant Hypoxia in Humans
Published on: December 22, 2016
Pulse oximetry to detect hypoxemia during apnea: comparison of finger and ear probes
Peter Lindholm1, S Lesley Blogg, Mikael Gennser
1Section of Environmental Physiology, Department of Physiology and Pharmacology, Karolinska Institutet, Stockholm, Sweden. Peter.Lindholm@ki.se
Introduction:
When investigating apnea, for example in diving or altitude studies, hypoxemia is a variable that must be monitored to reduce the risk of hypoxic syncope. Pulse oximetry is a simple technique that measures arterial oxygen saturation (SpO2). As apnea induces a peripheral vasoconstriction, we hypothesized that it would be better to measure hypoxia using more centrally placed ear lobe oximetry probes rather than peripheral finger probes.
Methods:
Seven men were studied, ages 18-35. Two pulse oximeters were used, a Satlite Trans (Ox-1) and Ohmeda Biox (Ox-2), both with ear and finger probes. Subjects carried out a sub-maximal breath hold for 60 s while performing dynamic leg exercise on a cycle ergometer at 50 W. Subjects performed the maneuver six times in total, in a crossover design.
Results:
The Ox-1 finger probe showed 6.0 +/- 3.7% higher values than the ear-lobe probe at their respective nadirs. The Ox-2 probes differed in the same manner by 6.5 +/- 4.2%. The average delay between the nadir shown by the ear and finger probes was 15 s (+/- 3.5). When the ear-probes were at their nadir (SpO2 78 +/- 3.5%), the finger probes had considerably higher SpO2 levels (94.6 +/- 3.5%).
Discussion:
Apneic induced hypoxemia was monitored poorly by finger probe pulse oximetry. The delay in response may jeopardize safety, for example in breath-hold diving studies. Hypoxemia does not seem to be accurately reflected by finger measurements in situations where peripheral vasoconstriction may occur.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...