Related Experiment Video
Updated: Jul 12, 2026

Air Filter Devices Including Nonwoven Meshes of Electrospun Recombinant Spider Silk Proteins
Published on: May 8, 2013
Performance of membrane filters used for TEM analysis of asbestos
James S Webber1, Alex G Czuhanich, Laurie J Carhart
1New York State Department of Health, Wadsworth Center, Albany, New York 12201, USA. webber@wadsworth.org
Abstract:
This article presents findings related to characteristics of membrane filters that can affect the recovery of asbestos and the quality of preparations for transmission electron microscopy (TEM) analysis. Certain applications and preparation steps can lead to unacceptable performance of membrane filters used in analysis of asbestos by TEM. Unless substantial care is used in the collapsing of mixed-cellulose ester (MCE) filters with an acetone hot block, grid preparations can suffer and fiber recoveries can be compromised. Calibration of the etching depth of MCE filters, especially at differing locations in an asher's chamber, is critical for reliable fiber recovery. Excessive etching of MCE filters with aerosol-deposited asbestos can lead to loss of short fibers, while insufficient etching of MCE filters with aqueous-deposited asbestos can, paradoxically, also lead to loss of short fibers. Interlaboratory precision on MCE filters is improved by aerosol-deposited asbestos, as opposed to aqueous deposition. In comparison, straightforward preparation, improved solvents, and reduced contamination make PC filters an increasingly acceptable alternative. Variations in the geometric configuration during application of carbon films can lead to fiber loss and unacceptable grid quality for either type of filter.