Related Experiment Video
Updated: Jul 12, 2026

Photoelectron Imaging of Anions Illustrated by 310 Nm Detachment of F−
Published on: July 27, 2018
Combining high mass resolution and velocity imaging in a time-of-flight ion spectrometer using pulsed fields and an
G Prümper1, H Fukuzawa, T Lischke
1Institute of Multidisciplinary Research for Advanced Materials, Tohoku University, Sendai, Japan. pruemper@tagen.tohoku.ac.jp
Abstract:
We describe a momentum resolving time-of-flight ion mass spectrometer that combines a high mass resolution, a velocity focusing condition for improved momentum resolution, and field-free conditions in the source region for high resolution electron detection. It is used in electron-ion coincidence experiments to record multiple ionic fragments produced in breakup reactions of small to medium sized molecules, such as F(3)SiCH(2)CH(2)Si(CH(3))(3). These breakup reactions are caused by soft x rays or intense laser fields. The ion spectrometer uses pulsed extraction fields, an electrostatic lens, and a delay line detector to resolve the position. Additionally, we describe a simple analytical method for calculating the momentum from the measured hit position and the time of flight of the ions.