Published on: May 31, 2016
Identification of calcifications in cardiac valves by near infrared Raman spectroscopy
Rick Rocha1, Antonio Balbin Villaverde, Carlos Augusto Pasqualucci
1Instituto de Pesquisa & Desenvolvimento (IP&D), Universidade do Vale do Paraíba - UNIVAP, Urbanova, São José dos Campos, SP, Brazil.
Objective:
The objective of this work was to detect calcification in cardiac valves using near infrared Raman spectroscopy (NIRS). A Ti:sapphire laser pumped by an argon-ion laser operating at a wavelength of 830 nm was used for excitation of the valve samples, and Raman emission was detected by an optical spectrometer with a liquid nitrogen-cooled CCD detector.
Background:
Cardiac valves are subjected to highly repetitive mechanical stresses, due to their over 40 million cardiac cycles per year. These structures may suffer cumulative lesions, complicated by the deposition of calcium phosphate, which can lead to clinically significant diseases. NIRS can provide important information about biological tissue composition and has been used for diagnosis of some types of human pathology.
Methods:
Samples of normal and pathologic tissues 5 mm in size were analyzed.
Results:
It was observed that the Raman spectrum of calcified cardiac valves presented different behavior when compared with normal valves. Differences were observed at the intensity of 960, 1,260, 1,452, and 1,660 cm(1) peaks.
Conclusions:
These results suggest that this technique could be used to detect calcium phosphate mineral deposition in cardiac valves.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...