Related Experiment Video
Updated: Jul 11, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Compensators for dose and scatter management in cone-beam computed tomography
S A Graham1, D J Moseley, J H Siewerdsen
1Ontario Cancer Institute, Princess Margaret Hospital, Department of Medical Biophysics, University of Toronto, Ontario, M5G 2M9, Canada.
Abstract:
The ability of compensators (e.g., bow-tie filters) designed for kV cone-beam computed tomography (CT) to reduce both scatter reaching the detector and dose to the patient is investigated. Scattered x rays reaching the detector are widely recognized as one of the most significant challenges to cone-beam CT imaging performance. With cone-beam CT gaining popularity as a method of guiding treatments in radiation therapy, any methods that have the potential to reduce the dose to patients and/or improve image quality should be investigated. Simple compensators with a design that could realistically be implemented on a cone-beam CT imaging system have been constructed to determine the magnitude of reduction of scatter and/or dose for various cone-beam CT imaging conditions. Depending on the situation, the compensators were shown to reduce x-ray scatter at the detector and dose to the patient by more than a factor of 2. Further optimization of the compensators is a possibility to achieve greater reductions in both scatter and dose.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
