Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
[Comparison between two methods for hemodynamic measurement: thermodilution and oesophageal doppler]
Roberto Manara Victorio Ferreira1, José Luiz Gomes do Amaral, Jorge Luís dos Santos Valiatti
1Centro de Ensino e Treinamento, Unifesp. rmanara@yahoo.com
Oesophageal Doppler (OD) offers accurate hemodynamic monitoring, comparable to the gold standard thermodilution (TD) method. This less invasive technique easily measures cardiac output (CO) in diverse clinical settings.
Area of Science:
- Cardiovascular Physiology
- Medical Device Technology
- Critical Care Medicine
Context:
- Hemodynamic monitoring is crucial for managing critically ill patients.
- Thermodilution (TD) is the established gold standard but is invasive.
- Oesophageal Doppler (OD) presents a less invasive alternative for hemodynamic assessment.
Purpose:
- To evaluate the accuracy of Oesophageal Doppler (OD) in measuring cardiac output (CO).
- To compare OD measurements against the gold standard thermodilution (TD) method.
- To assess the feasibility and ease of use of OD in clinical practice.
Summary:
- Simultaneous measurements of cardiac output (CO) were performed using both OD and TD in 10 patients across various clinical conditions.
- High correlation coefficients (0.8 and 0.99) were observed between OD and TD, indicating no statistically significant difference.
- OD measurements were easily obtained, demonstrating comparable accuracy to TD.
Impact:
- OD provides a reliable and less invasive method for hemodynamic monitoring.
- This finding supports the wider adoption of OD in clinical settings, potentially improving patient care.
- OD facilitates accurate cardiac output assessment, aiding in timely clinical decisions.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pipe Flowrate Measurement
The orifice meter is a simple,...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure