Reconstruction of rapidly acquired Germanium-68 transmission scans for cardiac PET attenuation correction
Bai-Ling Hsu1, James A Case, Kevin W Moser
1Cardiovascular Imaging Technologies, LLC, Kansas City, MO 64111, USA. bhsu@cvit.com
Background:
Transmission (TX) scan time by use of radionuclide sources for cardiac positron emission tomography prolong imaging and increase the likelihood of patient motion artifacts. A reconstruction algorithm combining ordered-subsets expectation maximization with a Bayesian prior was developed and applied to rapid Germanium-68 (Ge-68) TX scans.
Methods And Results:
A cardiac phantom with Fluorine-18 (Fl-18) was used to determine a minimal count threshold for Ge-68 TX scanning. Images were acquired over a count range from 2.5 x 10(6) to 8 x 10(7) and for a high-count scan of 1.6 x 10(9) counts to study reconstruction parameters and to determine the minimum TX count threshold. The method was compared against clinical 4-minute TX scans in ten Rubidium-82 (Rb-82) rest/stress myocardial perfusion studies (body mass index, 30 +/- 4 kg/m(2)). The minimal count threshold was 20 x 10(6), and the mean scan time for the Rb-82 studies was 70.5 +/- 3.4 seconds. More than 90% of the segmental scores computed from images acquired via rapid TX scans differed by less than 5% from those obtained with 4-minute TX scans. The mean differences in perfusion scores between the rapid and 4-minute TX scans were 0.46% (95% confidence interval, -1.84% to 0.93%) at rest and 0.39% (95% confidence interval, -1.84% to 1.07%) at stress, demonstrating equivalency of the rapid and 4-minute scans.
Conclusions:
Ordered-subsets expectation maximization with a Bayesian prior accurately and efficiently reconstructs rapidly acquired Ge-68 TX scans for Rb-82 myocardial perfusion positron emission tomography studies.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...