Determination of plasma tenofovir concentrations using a conventional LC-MS method
Masaaki Takahashi1, Yuichi Kudaka, Naoya Okumura
1Department of Pharmacy, National Hospital Organization Nagoya Medical Center (Tokai Area Central Hospital for AIDS Treatment and Research), Naka-ku, Nagoya, Aichi 460-0001, Japan.
Biological & Pharmaceutical Bulletin
|September 11, 2007
Summary
Quantifying tenofovir, a crucial HIV medication, in human plasma is now easier. A new liquid chromatography-mass spectrometry (LC-MS) method offers a precise and rapid way to measure tenofovir levels.
Area of Science:
- Pharmacology
- Analytical Chemistry
- Biomedical Science
Background:
- Tenofovir is a vital nucleoside reverse transcriptase inhibitor used daily for HIV treatment.
- Increased tenofovir usage necessitates accurate quantification in human plasma.
- Traditional HPLC methods face challenges due to interfering peaks, limiting tenofovir detection.
Purpose of the Study:
- To develop and validate a rapid and conventional LC-MS method for tenofovir quantification in human plasma.
- To establish a reliable analytical technique for therapeutic drug monitoring of tenofovir.
Main Methods:
- Development of a liquid chromatography-mass spectrometry (LC-MS) assay.
- Validation of the method for precision and accuracy in inter- and intraday analyses.
- Analysis of tenofovir in human plasma samples within a concentration range of 0.019-1.567 µg/mL.
Main Results:
- The LC-MS method demonstrated linearity across the tested concentration range.
- Average accuracy ranged from 95.9% to 100.7%.
- Inter- and intraday relative standard deviations were below 11.6%, with tenofovir recovery exceeding 80.2%.
Conclusions:
- A novel, rapid, and conventional LC-MS method for tenofovir determination in human plasma has been successfully developed.
- This validated method provides accurate and precise quantification, addressing limitations of previous analytical techniques.
- The established LC-MS assay is suitable for routine clinical and research applications involving tenofovir plasma level monitoring.