Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Continuous cardiac output measurement: arterial pressure analysis versus thermodilution technique during cardiac
S Lorsomradee1, S R Lorsomradee, S Cromheecke
1Department of Anaesthesia, Chiangmai University Hospital, Thailand.
Abstract:
This study compared cardiac output measured with an arterial pressure-based cardiac output measurement system and a thermodilution cardiac output measurement system. We studied 36 patients undergoing cardiac surgery with cardiopulmonary bypass. Simultaneous arterial pressure-based and thermodilution cardiac output measurements were compared before and after cardiopulmonary bypass, and after phenylephrine administration. Bland-Altman analysis showed good overall agreement between the two methods. Bias (limits of agreement) before and after cardiopulmonary bypass were - 0.21 (- 2.97-2.55) lxmin(-1) and 0.01 (- 3.79-3.81) lxmin(-1), respectively. Phenylephrine administration decreased thermodilution cardiac output by a mean (SD) of 11 (16)% and increased arterial pressure-based cardiac output by 55 (34)%. We conclude that arterial pressure-based cardiac output and thermodilution cardiac output measurement systems yield comparable results during cardiac surgery with cardiopulmonary bypass. However, after phenylephrine administration, the two measurement systems provided opposing results.
More Related Videos
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome (ARDS)
Published on: April 7, 2021
08:12Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)