Repeatability of central aortic blood pressures measured non-invasively using radial artery applanation tonometry and
Mike Crilly1, Christoph Coch, Margaret Bruce
1Department of Public Health, University of Aberdeen Medical School, Polwarth Building at Foresterhill, Aberdeen, UK. mike.crilly@abdn.ac.uk
Objective:
To estimate the repeatability of radial pulse wave analysis (PWA) in measuring central systolic and diastolic blood pressures (cSBP/cDBP), pulse pressure (cPP), augmentation pressure (cAP) and pulse pressure amplification (PPA).
Methods:
After 15 min supine rest, 20 ambulant patients (aged 27-82 years; four female) underwent four SphygmoCor PWA measurements on a single occasion. Two nurses independently undertook two measurements in alternate order, blind to their colleague's measurements. Analysis was by Bland-Altman limits of agreement (LOA).
Results:
Heart rate and brachial blood pressure (BP) were stable during assessment. Based on the average of two PWA measurements between-observer differences (LOA, mean difference +/- 2SD) were small (cSBP 1.5 +/- 10.9 mmHg; cDBP 0.4 +/- 5.2 mmHg; cAP 0.5 +/- 4.5 mmHg; cPP 1.1 +/- 10.5 mmHg; PPA -0.5% +/- 5.6%). Between-observer differences were much greater for single/initial PWA measurement (cSBP 3.6 +/- 15.9 mmHg; cDBP 2.8 +/- 8.8 mmHg; cAP 0.7 +/- 5.8 mmHg; cPP 0.8 +/- 13.6 mmHg; PPA -1.2 +/- 9.4%). Within-observer LOA were very similar for both nurse A (cSBP -4.2 +/- 14.1 mmHg; cDBP -4.6 +/- 13.1 mmHg; cAP -0.4 +/- 4.4 mmHg; cPP 0.5 +/- 11.0 mmHg; PPA 0.7% +/- 9.0%) and nurse B (cSBP 0.0 +/- 12.1 mmHg; cDBP 0.2 +/- 8.5 mmHg; cAP -0.1 +/- 4.4 mmHg; cPP -0.2 +/- 11.9 mmHg; PPA -0.7% +/- 10.6%).
Conclusion:
Non-invasive assessment of central aortic pressures using PWA on a single occasion is highly repeatable in ambulant patients even when used by relatively inexperienced staff.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...