Evaluation of two applanation tonometers in cats
P E Miller1, J P Pickett, L J Majors
1Department of Surgical Science, School of Veterinary Medicine, University of Wisconsin-Madison 53706.
American Journal of Veterinary Research
|November 1, 1991
Summary
This study found that both the MacKay-Marg and Tono-Pen tonometers underestimate intraocular pressure (IOP) in cats. However, predictable formulas were derived to correct these IOP measurements for both devices.
Area of Science:
- Veterinary Ophthalmology
- Ophthalmology
- Animal Health
Background:
- Intraocular pressure (IOP) measurement is crucial for diagnosing and managing ocular conditions in cats.
- Applanation tonometry is a common method for IOP assessment, but device accuracy can vary.
- The MacKay-Marg and Tono-Pen are widely used tonometers, necessitating a comparison of their performance in feline eyes.
Purpose of the Study:
- To compare the accuracy of the MacKay-Marg and Tono-Pen applanation tonometers in feline eyes.
- To establish correction formulas for IOP measurements obtained from these tonometers in cats.
- To evaluate the influence of tonometer order on IOP readings in vivo.
Main Methods:
- In vitro comparison of MacKay-Marg and Tono-Pen against direct manometry using feline eyes.
- In vivo IOP measurements in clinically normal cats, with varying sequences of tonometer application.
- Statistical analysis to determine significant differences and develop predictive IOP correction formulas.
Main Results:
- Both tonometers significantly underestimated IOP compared to direct manometry (P < 0.001) with high correlation (r2 = 0.99).
- Predictive formulas were derived: Actual IOP = 1.36 x (MacKay-Marg) - 1.67 and Actual IOP = 1.37 x (Tono-Pen) + 0.8.
- In vivo, the Tono-Pen consistently underestimated IOP compared to the MacKay-Marg (P < 0.001), with readings being lower when Tono-Pen followed MacKay-Marg (P < 0.05).
Conclusions:
- The MacKay-Marg and Tono-Pen require IOP correction formulas for accurate assessment in cats.
- The order of tonometer use can influence Tono-Pen readings, suggesting potential for measurement bias.
- These findings provide essential data for improving the clinical accuracy of IOP measurements in feline patients.