Testing the immunity of active implantable medical devices to CW magnetic fields up to 1 MHz by an immersion method
Valentin Buzduga1, Donald M Witters, Jon P Casamento
1Scantek, Inc, Columbia, MD 21046, USA. buzdugav@yahoo.com
Abstract:
This paper presents a magnetic-field system and the method developed for testing the immunity of the active implantable medical devices to continuous-wave magnetic fields in the frequency range up to 1 MHz. The system is able to produce magnetic fields of 150 A/m for frequencies up to 100 kHz and strengths decreasing as 1/f between 100 kHz and 1 MHz, with uniformity of the field within +/-2.5% in the volume for tests. To simulate human tissue, the medical device, together with its leads, is placed on a plastic grid in a saline tank that is introduced in the magnetic field of the induction coil. This paper offers an alternative for the injection voltage methods provided in the actual standards for assessing the protection of the implantable medical devices from the effects of the magnetic fields up to 1 MHz. This paper presents the equipment and signals used, the test procedure, and results from the preliminary tests performed at the Food and Drug Administration-Center for Devices and Radiological Health on implantable pacemakers and neurostimulators. The new system and test method are useful for the EMC research on the implantable medical devices.