Review of the first 1502 cases of ECG-ST waveform analysis during labour in a teaching hospital
V Doria1, A T Papageorghiou, A Gustafsson
1Clinical Developmental Sciences, Section of Obstetrics and Gynaecology, St George's University of London, London, UK.
Objective:
To assess the impact of introduction of the STAN monitoring system.
Study Design:
Prospective observational study.
Setting:
Tertiary referral labour ward, St George's Hospital, London.
Population:
High-risk term pregnancies.
Methods:
We report all consecutive cases of intrapartum monitoring using the STAN S 21 fetal heart monitor. Cases with adverse neonatal outcome were evaluated in relation to the ST waveform analysis and cardiotocography (CTG).
Main Outcome Measures:
Cord artery metabolic acidosis, neonatal encephalopathy (NNE) and reasons behind cases with poor outcome.
Results:
Between 2002 and 2005, there were 1502 women monitored by STAN. Based on combined STAN analysis in the 1502 women, action was indicated in 358 women (23.8%), while in 1108 women (73.8%) no action was indicated. Traces were not interpretable in 36 women (2.4%). Of the 836 cases (55.7%) where cord blood gases were available, there were 23 cases (2.8%) of metabolic acidosis and 16 of these (70%) were identified by STAN. Overall, there were 14 cases of NNE monitored by STAN. Retrospective analysis of these highlights human errors, such as poor CTG interpretation, delay in taking appropriate action and not following the guidelines.
Conclusions:
Our experience suggests the need for more intense training on interpretation of CTG and strict adherence to guidelines.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Dysrhythmias V: Evaluating Dysrhythmias
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...