[Conversion coefficients for estimation of effective dose to patients from dose area product during fluoroscopy x-ray
1Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Leipzig AöR. gosd@medizin.uni-leipzig.de
Purpose:
The effective dose, which is related to the stochastic radiation risk, cannot be measured directly. Therefore, conversion coefficients for estimating the effective dose for patients were calculated for fluoroscopy X-ray examinations from the dose area product which can be easily measured.
Materials And Methods:
Conversion coefficients were calculated using the PC-based Monte Carlo program PCXMC in software version 1.5.1. Conversion coefficients were determined for tube voltages between 60 kV and 110 kV, for additional cupper filtration up to 0.9 mm, for different projections (a. p., p. a. and lateral) and X-ray field size (between 20 cm x 20 cm and 40 cm x 40 cm in image plane) and field position (from head to leg).
Results:
Tube voltage, filtration and field position have a significant effect on conversion coefficients. Conversion coefficients increase as the tube voltage and filtration increase. Depending on the radiation spectrum used, the conversion coefficients may differ up to a factor of 4 for the same X-ray examination. The highest conversion coefficients were found for irradiation of the body region followed by the head and leg region. The effect of the field size and projection is moderate. Conversion coefficients for a. p. projection are higher than those for p. a. and lateral projections.
Conclusion:
Effective dose can be easily estimated by multiplying relevant conversion coefficients by the dose area product. Using conversion coefficients from the literature that do not take additional cupper filtration into consideration can underestimate the effective dose by a factor or up to 2 or more depending on the filtration used.
Related Concept Videos
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant