Assessment of the residual error in soft tissue setup in patients undergoing partial breast irradiation: results of a
Toyosi Fatunase1, Zhiheng Wang, Sua Yoo
1Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27710, USA.
Purpose:
On-board cone-beam computed tomography (CBCT) provides soft tissue information that may improve setup accuracy in patients undergoing accelerated partial breast irradiation (APBI). We used CBCT to assess the residual error in soft tissue after two-dimensional kV/MV alignment based on bony anatomy. We also assessed the dosimetric impact of this error.
Methods And Materials:
Ten patients undergoing APBI were studied as part of an institutional review board-approved prospective trial. Patients were aligned based on skin/cradle marks plus orthogonal kV/MV images registered based on bony landmarks to digitally reconstructed radiographs from the planning CT. A subsequent CBCT was registered to the planning CT using soft tissue information. This "residual error" and its dosimetric impact was measured.
Results:
The root-mean-square of the residual error was 3, 4, and 4 mm, in the right-left, anterior-posterior, and superior-inferior directions, respectively. The average vector sum was 6+/-2 mm. Average reductions in mean dose to the lumpectomy cavity, clinical target volume (CTV), and planning target volume were 0.1%, 0.4%, and 1%, respectively. The mean difference in the clinical target and planning target volumes that received 95% of the prescribed dose (V95) were 1% and 4%.
Conclusions:
In this initial study with a modest number of patients, the residual error in soft tissue was typically <5 mm, and with the field margins used, the resultant dosimetric consequences were modest. In patients immobilized in a customized cradle, setup using orthogonal kV images thus appears accurate and reproducible. The CBCT technique may have particular utility in patients with larger breast volumes or breast deformations. Further studies involving larger numbers of patients are needed to further assess the utility of CBCT.
More Related Videos
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography