Influence of inversion pulse type in assessing lung-oxygen-enhancement by centrically-reordered non-slice-selective
Michael Puderbach1, Yoshiharu Ohno, Hideaki Kawamitsu
1Department of Radiology, Deutsches Krebsforschungszentrum, Heidelberg, Germany.
Purpose:
To demonstrate the influence of inversion pulse type and inversion time for assessment of oxygen-enhancement on centrically-reordered non-slice-selective inversion-recovery (IR) half-Fourier single-shot turbo spin-echo (HASTE) sequence.
Material And Methods:
Phantoms with and without 100% oxygen and three healthy volunteers were studied with two-dimensional (2D) centrically-reordered non-slice selective IR-HASTE sequence with either composite or block inversion-recovery pulse at increasing inversion times from 200 to 1800 msec. Signal-to-noise ratios (SNRs) of phantom, real signal differences, and relative enhancement ratios of lung parenchyma between oxygen-enhanced and non-oxygen-enhanced MR images on composite and block pulse type were statistically compared at each TI.
Results:
SNRs at TIs of 200 and 400 msec using the composite inversion pulse type were significantly lower than those with the block inversion pulse in the in vivo study (P < 0.05), although no significant differences were observed in the phantom study and in the in vivo study at inversion times greater than or equal to 600 msec. Real signal intensity (SI) differences at 400 and 600 msec of the composite inversion pulse type were significantly higher than those with the block inversion pulse type (P < 0.05). Relative enhancement ratio at 800 msec with the composite inversion pulse were significantly lower than that with the block inversion pulse (P < 0.05).
Conclusion:
IR pulse type and inversion time have influence on assessment of oxygen-enhancement by centrically-reordered non-slice-selective IR-HASTE sequence.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Nuclear Overhauser Enhancement (NOE)
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Imaging Studies for Cardiovascular System IV: CMRI