Parallel imaging of knee cartilage at 3 Tesla
Jin Zuo1, Xiaojuan Li, Suchandrima Banerjee
1Musculoskeletal and Quantitative Imaging Research (MQIR), Department of Radiology, University of California, San Francisco, San Francisco, California, USA.
Purpose:
To evaluate the feasibility and reproducibility of quantitative cartilage imaging with parallel imaging at 3T and to determine the impact of the acceleration factor (AF) on morphological and relaxation measurements.
Materials And Methods:
An eight-channel phased-array knee coil was employed for conventional and parallel imaging on a 3T scanner. The imaging protocol consisted of a T2-weighted fast spin echo (FSE), a 3D-spoiled gradient echo (SPGR), a custom 3D-SPGR T1rho, and a 3D-SPGR T2 sequence. Parallel imaging was performed with an array spatial sensitivity technique (ASSET). The left knees of six healthy volunteers were scanned with both conventional and parallel imaging (AF = 2).
Results:
Morphological parameters and relaxation maps from parallel imaging methods (AF = 2) showed comparable results with conventional method. The intraclass correlation coefficient (ICC) of the two methods for cartilage volume, mean cartilage thickness, T1rho, and T2 were 0.999, 0.977, 0.964, and 0.969, respectively, while demonstrating excellent reproducibility. No significant measurement differences were found when AF reached 3 despite the low signal-to-noise ratio (SNR).
Conclusion:
The study demonstrated that parallel imaging can be applied to current knee cartilage quantification at AF = 2 without degrading measurement accuracy with good reproducibility while effectively reducing scan time. Shorter imaging times can be achieved with higher AF at the cost of SNR.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...