Related Experiment Video
Updated: Jul 11, 2026

Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Effects of left ventricular unloading by Impella recover LP2.5 on coronary hemodynamics
Maurice Remmelink1, Krischan D Sjauw, José P S Henriques
1Department of Cardiology, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands.
Objectives:
We studied the effects of LV unloading by the Impella on coronary hemodynamics by simultaneously measuring intracoronary pressure and flow and the derived parameters fractional flow reserve (FFR), coronary flow velocity reserve (CFVR), and coronary microvascular resistance (MR).
Background:
Patients with compromised left ventricular (LV) function undergoing high-risk percutaneous coronary intervention (PCI) may benefit from LV unloading. Limited information is available on the effects of LV unloading on coronary hemodynamics.
Methods:
Eleven patients (mean LV ejection fraction of 35 +/- 11%) underwent PCI during LV support by the LV unloading device (Impella Recover LP2.5). Intracoronary measurements were performed in a nonstenotic coronary artery after the PCI, before and after adenosine-induced hyperemia at four different support levels (0-2.5 L/min).
Results:
Aortic and coronary pressure increased with increasing support levels, whereas FFR remained unchanged. Baseline flow velocity remained unchanged, while hyperemic flow velocity and CFVR increased significantly with increasing support levels (61 +/- 24 to 72 +/- 27 cm/sec, P = 0.001 and 1.88 +/- 0.52 to 2.34 +/- 0.63, P < 0.001 respectively). The difference between baseline MR and hyperemic MR significantly increased with increasing support levels (1.28 +/- 1.32 to 1.89 +/- 1.43 mm Hg cm(-1) sec, P = 0.005).
Conclusions:
Unloading of the LV by the Impella increased aortic and intracoronary pressure, hyperemic flow velocity and CFVR, and decreased MR. The Impella-induced increase in coronary flow, probably results from both an increased perfusion pressure and a decreased LV volume-related intramyocardial resistance.
More Related Videos
08:49Cardiac Loading using Passive Left Atrial Pressurization and Passive Afterload for Graft Assessment
Published on: August 2, 2024
05:25Surgical Porcine Model of Chronic Myocardial Ischemia Treated by Exosome-laden Collagen Patch and Off-pump Coronary Artery Bypass Graft
Published on: September 15, 2023