Related Experiment Video
Updated: Jul 11, 2026

Fabrication and Operation of Acoustofluidic Devices Supporting Bulk Acoustic Standing Waves for Sheathless Focusing of Particles
Published on: March 6, 2016
Piezoelectric pumping in flow analysis: Application to the spectrophotometric determination of gabapentin
Marta F T Ribeiro1, João L M Santos, José L F C Lima
1REQUIMTE, Faculdade de Farmácia da Universidade do Porto, Rua Aníbal Cunha 164, 4099-030 Porto, Portugal.
Abstract:
Solutions propelling devices are fundamental components of a flow-based analytical manifold. In this work different manifold configurations were implemented to evaluate the performance of multiple piezoelectric micro-pumps used as solutions insertion and propulsion devices. The micro-pumps are piezo-actuated micro-diaphragm pumps with passive check valves characterised by a small compact size and low power demands, and are able to produce reproducible flow rates of up to 4 mL min(-1). The flow rate is controlled by the frequency of the piezoelectric actuator (up to 20 Hz). As an additional feature, piezoelectric micro-pumps actuation generates a pulsed flowing stream that ensures a faster sample/reagent mixing contributing to improved reaction development. The developed flow approach was assessed in the spectrophotometric determination of gabapentin in pharmaceutical preparations upon reaction with 1,2-naphthoquinone-4-sulfonate in alkaline medium. Distinct flow manifold configurations were designed for achievement of different solutions management. Linear calibrations plots for gabapentin concentrations of up to 150 mg L(-1), with relative standard deviations of less than 1.50% (n=10) and a sample throughput of about 28 determinations per hour, were obtained.