Related Experiment Video
Updated: Jul 11, 2026

A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
End-column chemiluminescence detection for pressurized capillary electrochromatographic analysis of norepinephrine
Zian Lin1, Xiaoping Wu, Xucong Lin
1Department of Chemistry, Fuzhou University, Fuzhou 350002, China.
Abstract:
A simple and convenient end-column chemiluminescence (CL) detection coupled to pressurized capillary electrochromatography (pCEC) was described. Luminol and N-(4-aminobutyl)-N-ethylisoluminol (ABEI) were adopted as mode compounds to evaluate the feasibility of end-column reactor. Detailed analysis of ABEI revealed that the high sensitivity could be obtained with the reactor. Furthermore, determination of norepinephrine (NE) and epinephrine (EP), which were labeled with ABEI, was accomplished by using the end-column pCEC-CL detection based on ABEI-potassium ferricyanide-alkaline medium CL reaction system. Under the optimum conditions, the detection limit (S/N=3) of NE and EP was 0.08 microM and 0.06 microM, respectively. The proposed method has also been successfully applied to the analysis of adrenaline hydrochloride injection sample.
More Related Videos
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
High-Performance Liquid Chromatography: Types of Detectors
Capillary Electrophoresis: Instrumentation
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Detectors-II